Amplitude Detection
- Set the controls as follows:
- Channel 3 Image Data Type: Amplitude
- Z range: 1 nm
- Line direction: Retrace.
- Change Interleave Mode to Lift to invoke LiftMode.
- Set the Channel 3 Scan Line to Interleave to display the interleaved data.
NOTE: This can only be done after Interleave mode is enabled. A magnetic force gradient image similar to that shown on the right of figure 1 (below) should appear as the Channel 3 image. The alternating dark and light stripes represent the recorded magnetic information, signifying a varying resonant frequency and magnetic force gradient on the tip.
Figure 1: Topographic (left) and Magnetic Force Gradient (right) AFM images of metal-evaporated video tape.
HINT: Keep the Setpoint as large as possible while maintaining image quality. Wider scans (> 25 µm) will reveal separate tracks in which the magnetic stripes are at different angles.
www.bruker.com
|
Bruker Corporation |
www.brukerafmprobes.com
|
112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/
|
Santa Barbara, CA 93117 |
|
|
|
Customer Support: (800) 873-9750 |
|
Copyright 2010, 2011. All Rights Reserved. |
Open topic with navigation